Page 49 - MERC Flip Template

Basic HTML Version

Appendix B
5/20/2009
6
3.48.
Method Detection Limit (MDL)
- The minimum concentration of an analyte
that can be identified, measured, and reported with 98% confidence that the analyte
concentration is greater than zero.
3.49.
Mixing Volume
- Spherical bottle in which sample gases become
homogenous.
3.50.
Mother Board
- The main printed circuit board. All CE 440 power supplies
are located here.
3.51.
Must -
Denotes a requirement that must be met. (Random House College
Dictionary)
3.52.
Precision -
The degree to which a set of observations or measurements of the
same property, obtained under similar conditions, conform to themselves; a data
quality indicator. Precision is usually expressed as standard deviation, variance or
range, in either absolute or relative terms. (NELAC)
3.53.
Preservation –
Refrigeration, freezing and/or reagents added at the time of
sample collection (or later) to maintain the chemical and or biological integrity of the
sample.
3.54.
Pressure Transducer
- Used to check for leaks in the system and to monitor
pressure in the mixing volume.
3.55.
P Valve
- The valve on the injector box of the horizontal auto-injector (HA)
used to automatically purge the box.
3.56.
Profile
- Generated by the bridge signal. Used to help determine if a leak or
malfunction occurs in the system.
3.57.
Quality Control Sample (QCS)
- A sample of analytes of known and
certified concentrations. The QCS is obtained from a source external to the
laboratory and different from the source of calibration standards. It is used to check
laboratory performance with externally prepared test materials.
3.58.
Reduction Tube
- Quartz tube with reduced copper that removes excess
oxygen from the sample gas and reduces oxides of nitrogen to free nitrogen.
3.59.
Response Factor (RF)
- The ratio of the response of the instrument to a
known amount of analyte.
3.60.
Run
- One sample analysis from start to finish, including printout.
3.61.
Run Cycle
- Typically a day or half day of operation - the entire analytical
sequence of runs from the first run to the last run on the Sample Wheel.
3.62.
Sample Volume
- Tube where sample gas is exhausted from the mixing
volume prior to entering the detector.
3.63.
Sample Wheel
Sample holding device which contains up to 64 blanks,
standards and samples. One wheel equals roughly 6 hours of run time, which is
called the Run Cycle.
3.64.
Scrubber
- Removes water and CO
2
from the gas supplies.
3.65.
Sediment (or Soil) Sample
- A fluvial, sand, or humic sample matrix exposed
to a marine, estuarine or fresh water environment.
3.66.
Sensitivity -
The capability of a test method or instrument to discriminate
between measurement responses representing different levels (concentrations) of a
variable of interest.
3.67.
Shall -
Denotes a requirement that is mandatory whenever the criterion for
conformance with the specification requires that there be no deviation. (ANSI)