Page 66 - MERC Flip Template

Basic HTML Version

Appendix B
5/20/2009
23
17.2.3.4.H/C, N/C, C/H, C/N Ratio
17.2.3.5.C/C, H/H, N/N, O/O, S/S Ratio
17.2.3.6.Empirical Formula
17.2.4.
Customize Pull-Down Menu (Customizing software)
17.2.4.1.Set parameters
17.2.4.2.Users
17.2.4.3.Edit Standards (names, weights, percents)
17.2.4.4.Create Report Format
17.2.4.5.Change Infinite Run Counter
17.2.4.6.Set Automation Type
17.2.5.
Help
17.3.
Run Pull-Down Menu
17.3.1.
Select “Carbon, Hydrogen, Nitrogen Run”
17.3.2.
Select “Yes” for a new run
17.3.3.
Enter message for this run series
17.3.3.1.Check “Enter the Ks and Blanks automatically”.
17.3.3.2.Enter date followed by AM or PM as appropriate
17.3.3.3.Press “Enter Data”
17.3.4.
Sample Entry Screen
17.3.4.1.Enter Weight (µg)
17.3.4.1.1. When entering the weight of the sample press
[ENTER] to use the present weight or enter a
new weight. If a weight of zero [0] is entered
then the ID is assumed to be a blank. If a
weight of 100 has been entered the results will
be reported in micrograms (µg). When
analyzing aqueous samples, enter the volume
filtered/10 as the weight. The results will be
reported in ug/l. When analyzing sediment
samples or weighed QC samples, enter the
weight in ug. The result will be reported in %.
17.3.4.2.Enter Sample ID
17.3.4.2.1. Enter the sample ID as either STD1, blank, or
any other text. If STD is entered as the first
three letters, then Ks will be calculated on the
result report. If blank is entered, then blanks
will be calculated. If a weight of 100 has been
entered, the results will be reported in
micrograms (µg). If a “weight” of volume
filtered/10 has been entered, the results will be
in ug/l. If a weight of ug has been entered, the
result will be reported in %.