Appendix C
            
            
              
                5/20/2009
              
            
            
              15
            
            
              11.3.6
            
            
              The furnace automatically turns on when connected.  It takes
            
            
              20-30 minutes to come to temperature.
            
            
              11.3.7
            
            
              To view instrument conditions, click on View /Background
            
            
              Monitor.  When all conditions are OK, the system is ready to
            
            
              run.
            
            
              11.3.8
            
            
              Loading Samples: Decide on standard curve.  I use 2 curves:
            
            
              Low — 0-10 ppm ---- DHOH, 2.0, 5.0, 10.0 as calibration
            
            
              standards, and adding 0.5, 1.0, and 7.5 as well for stds as
            
            
              samples.     60 µL injection.                                        High-----
            
            
              0-20 ppm ----- DHOH, 5.0, 10.0, 20.0 as calibrating STDs and
            
            
              adding 1.0, 2.0, & 15.0 as STDs as samples.       30 µL
            
            
              injection.  The instrument allows up to a 4 point curve.
            
            
              11.3.9
            
            
              Use the large sample vials for the curve stds.   Examples of
            
            
              sample protocol can be found in the data sheet notebook.
            
            
              11.3.10 I always load several DHOH samples (at least 3) as
            
            
              conditioning samples
            
            
              11.3.11 Fill std vials ~ 1/3 to 1/2 full.
            
            
              11.3.12 Fill sample vials to within 3/4 - 1" of the top.
            
            
              11.3.13 Data sheets are found in the LOTUS TOC5000 folder on the
            
            
              desktop, labeled TOC datasheet, high, low std.  Print as many
            
            
              as needed.  (Low STD set to print 5; high STD prints 2, and
            
            
              plain datasheet prints 1.)
            
            
              11.3.14 Acidify the samples by adding 100 µl of 2N HCl.  The
            
            
              standard curve files are set up to automatically add acid to the
            
            
              standard cups.  Make sure a standard cup filled with 2 N HCl
            
            
              is in position S8.
            
            
              11.3.15  When the turntable is loaded, place in the ASI.  Line up pins
            
            
              to align properly.  Put top on lining up arrows carefully.
            
            
              Sample needles will Z if the top is not properly aligned.
            
            
              11.3.16 Close the Background Monitor and a blank sample table will
            
            
              be opened.
            
            
              11.3.17 Enter sample table by clicking on Edit/Insert Std. Click on
            
            
              CONDBLK.CAL.   Then click on Edit/Autogenerate.  In the
            
            
              method name slot, type CONDBLK.MET.  Using the TAB,
            
            
              toggle through to fill in the blanks.  This is an NPOC file.  I
            
            
              usually put conditioning DHOH vials in positions 1-3.  The
            
            
              standard curve is next.  Several curves are already stored.
            
            
              Click on Edit/ insert STD again to choose the proper curve.
            
            
              LOSTD10 is a 0-10 ppm curve with DHOH, 2.0, 5.0, & 10 in
            
            
              positions S1, 2, 3, 4.  HISTD20 is a 0-20 ppm curve with
            
            
              DHOH, 5.0, 10.0, 20.0 in positions S1, 2, 3, 4.
            
            
              11.3.18 Click on Edit/ Insert Sample next.  Most likely this is a new
            
            
              file, so click on NEW.  You will then be prompted enter the
            
            
              conditions of the samples to be run.  Enter a file name (up to 8
            
            
              characters), and TAB to the next slot. Enter the sample
            
            
              information (ex. DNR Potomac 3/30/08).  TAB again to enter